CO- 5.6.0 TROUBLESHOOTING UNSTABLE CONDUCTIVITY
Proceed through the following SEVEN STEPS until a problem is located:
1. ENSURE a ‘strong’ spring was NOT accidently installed at the output ports of the acid AND
bicarbonate pump. BOTH pumps get ‘weak’ springs!
2. Compare both springs to new springs. Replace any that are shorter than a new spring.
3. Other procedures, in different Sections of the Guide, are performed next.
NOTE
this page and
procedure number (CO- 5.6.0) because you may prompted to return here.
4. If not already performed in THIS Troubleshooting session, BEFORE continuing to step #5, proceed to
page 535
, to perform
SECTION 19 - TESTING FOR A LEAKING BALANCING CHAMBER
5. If a bad balancing chamber diaphragm was not located in step #4, BEFORE continuing to step #6,
proceed to
page 139
, to perform
6. If the Hydrochamber is okay, per step #5, ENSURE Deaeration Motor brushes have been replaced at
8000 hour intervals AND Flow Motor brushes have been replaced approximately at 16,000 hour
intervals.
7. Assuming ALL procedures were performed thoroughly AND Conductivity remains unstable the
Troubleshooting Guide is unable to locate the problem.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...