2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 253
T- 1.5.51 5V HI MORE THAN 5.3 AND / OR 24 V HI MORE THAN 28.0
A) Turn the machine OFF!
B) Open the card cage and push down HARD on ALL circuit boards!
C) Using compressed air, clean the motherboard’s surface of excessive dust.
D) Using a flashlight, inspect the surface of the motherboard for corrosion or burning!
E) BEFORE going to part F, note this page number, then proceed to
page 14
to perform the
F) Return to HEAT DISINFECT then IMMEDIATELY call debug screen 10. If debug doesn’t appear press
‘Esc’ then call screen 10.
G) Allow thirty (30) minutes OR until if “TEMP OVER 95 DEGREES” reoccurs. THREE (3) possible
scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) “TEMP OVER 95 DEGREES” DOES NOT reoccur:
DO NOT continue!
2) IF 5V HI = 5.3 OR MORE AND / OR 24V = 28.0 OR MORE:
Continue to part H.
3)
IF “TEMP OVER 95 DEGREES” reoccured BUT 5V HI REMAINED less than 5.3 AND 24V HI
REMAINED less than 28.0:
Proceed to
page 246
, procedure number’s
SCENARIO
#3
.
H)
Read before performing!
With the machine off, swap in one of the listed components (see
COMPONENT LIST below) with known good THEN, in between, perform parts I and J to see if the new
component fixes the high voltage problem.
COMPONENT LIST: 1
) Power Logic Board
1
;
2)
Actuator-Test Board
1, 2
;
3)
Sensor Board
1, 3
;
4)
Functional Board
1, 2, 3
;
5)
Power Supply;
6)
Motherboard
1
.
1
To LOCATE these boards refer to
(page 9)
2
BEFORE performing parts I and J note this page number then proceed to
page 14
3
To prevent “Cond Offset Failure”
place the machine into
T and C Mode
(refer to
OPERATING
MODES
, page
Error! Bookmark not defined.
))
I) Return to HEAT DISINFECT THEN IMMEDIATELY call debug screen 10. If debug does not appear
press ‘Esc’ then call screen 10.
J) Allowing up to thirty (30) minutes. Based on 5V HI AND 24V HI,
TWO (2) possible scenarios:
1)
IF (and ONLY if) 5V HI = 5.3 OR MORE AND / OR 24V = 28.0 OR MORE:
Return to above
part H.
2) IF 5V HI REMAINS LESS THAN 5.3 AND 24V HI REMAINS LESS THAN 28.0:
The last
component ‘swapped’ in solved the problem.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...