2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 628
P- 1.0.0 MACHINE NEVER TURNS ON (FAN NEVER RUNS)
a)
IMPORTANT! Set your CALIBRATED volt meter to DC voltage (V
DC
)!
b)
Connect the meter’s black lead to chassis ground (see
(page 4)).
c) Per the
below, locate regulator
T1
at the TOP EDGE of the Power Logic Board. TWO (2)
measurements may be made at T1 depending on the first!
d)
MEASUREMENT 1:
From the REAR side of T1’s
TOP pin
(pin 1). More than 10.0 volts DC?
Yes
More than 10.0 volts! The main line fuses are okay. See procedure number
No Less than 10.0 volts! Proceed to
page 634
, procedure number
Figure 120 – Power Logic Board (T1)
P- 1.0.1 PIN 1 MORE THAN 10.0 VOLTS / MEASUREMENT 2
MEASUREMENT 2:
Per the Figure above, from the rear side of T1’s
BOTTOM pin
(pin 3). Between 4.0
and 6.0 volts DC?