Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
930
User Manual 1178.8520.02 ─ 08
[SENSe:]ESPectrum<sb>:SCENter
<Frequency>
This command defines the center frequency of the selected sub block in a Multi-SEM
measurement.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
Parameters:
<Frequency>
Frequency within the currently defined global span (see
Range:
1 to 3
*RST:
1
Default unit: Hz
Example:
ESP1:SCEN 1GHZ
Manual operation:
See
[SENSe:]ESPectrum<sb>:SCOunt
<Subblocks>
This command defines the number of sub blocks in the SEM measurement.
Suffix:
<sb>
.
irrelevant
Parameters:
<Subblocks>
Number of sub blocks in the SEM measurement.
Range:
1 to 8
*RST:
1
Example:
ESP:SCO 2
Manual operation:
See
12.5.6.4
Configuring a sweep list
The following commands define a sweep list for SEM measurements.
The sweep list cannot be configured using remote commands during an on-going
sweep operation.
See also:
●
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
.................................................................................. 931
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:RESolution
........................................... 931
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:VIDeo
.................................................. 932
[SENSe:]ESPectrum<sb>:RANGe<ri>:COUNt?
.................................................................932
Configuring and performing measurements