<RBW>
Defines the resolution bandwidth for a list entry.
<VBW>
Defines the measurement time for a list entry.
<MeasTime>
Defines the measurement time for a list entry.
Range:
1 µs to 16000 s
Default unit: s
<TriggerLevel>
Reserved for future use; currently: must be 0.
Example:
See
Chapter 12.5.14.2, "Example: performing list evaluation"
[SENSe:]LIST:POWer:SET
<State>, <State>, <State>, <TriggerSource>,
<TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level
after
sending this command.
Parameters:
<State>
ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST:
1
<State>
ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST:
0
<State>
ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST:
0
<TriggerSource>
EXTernal | EXT2 | EXT3 | IMMediate | IFPower | RFPower |
VIDeo
Selects a trigger source.
<TriggerSlope>
POSitive | NEGative
Selects the trigger slop.
<TriggerOffset>
Defines the trigger delay.
Range:
negative measurement time to 30 s
*RST:
0
Default unit: s