Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
175
User Manual 1178.8520.02 ─ 08
Predefined Standards ← Standard
Predefined standards contain the main measurement settings for standard measure-
ments. When such a standard is loaded, the required channel settings are automati-
cally set on the R&S
FSV/A. However, you can change the settings.
The predefined standards contain the following settings:
●
Channel bandwidths
●
Channel spacings
●
Detector
●
Trace Average setting
●
Resolution Bandwidth (RBW)
●
Weighting Filter
For details on the available standards, see
Chapter 5.3.9, "Reference: predefined CP/
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
User Standards ← Standard
Access
: "CP / ACLR Config" > "General Settings" tab > "Manage User Standards"
Note:
User standards created on an analyzer of the R&S FSP family are compatible to
the R&S
FSV/A. User standards created on an R&S
FSV/A, however, are not necessa-
rily compatible to the analyzers of the R&S FSP family and may not work there.
The following parameter definitions are saved in a user-defined standard:
●
Number of adjacent channels
●
Channel bandwidth of transmission (Tx), adjacent (Adj) and alternate (Alt) chan-
nels
●
Channel spacings
●
Weighting filters
●
Resolution bandwidth
●
Video bandwidth
●
Detector
●
ACLR limits and their state
●
"Sweep Time" and "Sweep Time" coupling
●
Trace and power mode
Save the current measurement settings as a user-defined standard, load a stored mea-
surement configuration, or delete an existing configuration file.
For details see
Chapter 5.3.6.4, "How to manage user-defined configurations"
Remote command:
To query all available standards:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
on page 867
To load a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
To save a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE
Channel power and adjacent-channel power (ACLR) measurement