![Rohde & Schwarz FSV3000 User Manual Download Page 1235](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431235.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1235
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:CONTrol:OFFSet
<Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Offset>
Numeric value.
The unit depends on the scale of the x-axis.
*RST:
0
Default unit: HZ
Manual operation:
See
CALCulate<n>:LIMit<li>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Distance>
Numeric value.
The unit depends on the scale of the x-axis.
Default unit: HZ
Manual operation:
See
CALCulate<n>:LIMit<li>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
Manual operation:
See
Analyzing measurements (basics)