![Rohde & Schwarz FSV3000 User Manual Download Page 934](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843934.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
934
User Manual 1178.8520.02 ─ 08
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the SEM sweep list covers, the R&S
FSV/A will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FSV/A will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Chapter 5.6.4.1, "Ranges and range settings"
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value. Note that the minimum frequency range of a
SEM range is 20
Hz.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the SEM sweep list covers, the R&S
FSV/A will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FSV/A will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Chapter 5.6.4.1, "Ranges and range settings"
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Configuring and performing measurements