Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
874
User Manual 1178.8520.02 ─ 08
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
<State>
This command turns the weighting filter for a transmission channel on and off.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
12.5.3.4
Selecting the reference channel
The following commands define the reference channel for relative ACLR measure-
ments.
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
....................................................... 874
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
................................................ 874
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
............................................. 875
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
This command sets the channel power as the reference for relative ACLR measure-
ments.
Example:
POW:ACH:REF:AUTO ONCE
Usage:
Event
Manual operation:
See
"Setting a fixed reference for Channel Power measure-
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
<RefChannel>
This command selects the reference channel for relative measurements.
You need at least one channel for the command to work.
Parameters:
<RefChannel>
MINimum | MAXimum | LHIGhest
MINimum
Transmission channel with the lowest power
MAXimum
Transmission channel with the highest power
LHIGhest
Lowest transmission channel for lower adjacent channels and
highest transmission channel for upper adjacent channels
Example:
POW:ACH:REF:TXCH:AUTO MAX
Selects the channel with the peak power as reference channel.
Configuring and performing measurements