Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
489
User Manual 1178.8520.02 ─ 08
Remote command:
TRIG:SOUR PSE
, see
SWE:EGAT:SOUR PSE
for gated triggering, see
Time ← Trigger Source
Triggers in a specified repetition interval.
See
Remote command:
TRIG:SOUR TIME
, see
Trigger Level
Defines the trigger level for the specified trigger source.
For gated measurements, this setting also defines the gate level.
For details on supported trigger levels, see the instrument data sheet.
For time triggers, the repetition interval is defined. See
Remote command:
TRIGger[:SEQuence]:LEVel:IFPower
TRIGger[:SEQuence]:LEVel:IQPower
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
TRIGger[:SEQuence]:LEVel:VIDeo
Repetition Interval
Defines the repetition interval for a time trigger.
The shortest interval is 1
μ
s. The granularity of the repetition interval is 1/256
MHz.
Set the repetition interval to the exact pulse period, burst length, frame length or other
repetitive signal characteristic. If the required interval cannot be set with the available
granularity, configure a multiple of the interval that can be set. Thus, the trigger
remains synchronized to the signal.
Remote command:
TRIGger[:SEQuence]:TIME:RINTerval
Drop-Out Time
Defines the time that the input signal must stay below the trigger level before triggering
again.
For more information on the drop-out time, see
Remote command:
Trigger Offset
Defines the time offset between the trigger event and the start of the sweep.
For more information, see
Trigger and gate configuration