![Rohde & Schwarz FSV3000 User Manual Download Page 444](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843444.webp)
Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
444
User Manual 1178.8520.02 ─ 08
Providing trigger signals as output is described in detail in
and providing trigger signals"
Chapter 6.2.11, "How to output a trig-
Trigger 1/2
The trigger input and output functionality depends on how the variable "Trigger Input/
Output" connectors are used.
"Trigger 1"
"Trigger 1" is input only.
"Trigger 2"
Defines the usage of the variable "Trigger Input/Output" connector on
the front panel
"Trigger 3"
Defines the usage of the variable "Trigger 3 Input/Output" connector
on the rear panel
"Input"
The signal at the connector is used as an external trigger source by
the R&S
FSV/A. Trigger input parameters are available in the "Trig-
ger" dialog box.
"Output"
The R&S
FSV/A sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Remote command:
OUTPut<up>:TRIGger<tp>:DIRection
Output Type ← Trigger 1/2
Type of signal to be sent to the output
"Output Off"
Deactivates the output. (Only for "Trigger 3", for which only output is
supported.)
"Device Trig-
gered"
(Default) Sends a trigger when the R&S
FSV/A triggers.
Data input and output