Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
946
User Manual 1178.8520.02 ─ 08
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
[SENSe:]ESPectrum<sb>:RRANge?
This command queries the reference range.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
Return values:
<RefRange>
Number of the current reference range.
Range:
1 to 30
Usage:
Query only
[SENSe:]ESPectrum<sb>:RTYPe
<Type>
This command defines the type of the power reference.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
Parameters:
<Type>
PEAK | CPOWer
PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST:
CPOWer
Manual operation:
See
12.5.6.6
Configuring the power classes
The following commands define the power classes for SEM measurements.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:LIMits
................................................................947
CALCulate<n>:LIMit<li>:ESPectrum<sb>:MODE
...............................................................948
CALCulate<n>:LIMit<li>:ESPectrum<sb>:VALue
............................................................... 948
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:COUNt
...........................................949
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:EXCLusive]
................................... 949
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
.................................. 950
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum
...................................... 950
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum
....................................... 951
Configuring and performing measurements