Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
936
User Manual 1178.8520.02 ─ 08
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the gain for a SEM range.
In case of high speed measurements, the level of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1 to 8
<ri>
1..30
Selects the measurement range.
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a SEM range on and off.
In case of high speed measurements, the state of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INSert
<Mode>
This command inserts a new SEM range and updates the range numbers accordingly.
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the SEM range.
Parameters:
<Mode>
AFTer | BEFore
AFTer
Inserts a range after the selected range.
BEFore
Inserts a range before the selected range.
Manual operation:
See
"Insert before Range / Insert after Range"
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt
<Level>
This command defines an absolute limit for a SEM range.
Configuring and performing measurements