![Rohde & Schwarz FSV3000 User Manual Download Page 1046](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431046.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1046
User Manual 1178.8520.02 ─ 08
This information is also indicated on the electronic paper display on the side panel of
the device.
Beware that if you change the network setting to DHCP = ON, the connection is abor-
ted and you must re-establish a connection to the frontend (see
]EFRontend<fe>:CONNection[:STATe]
Suffix:
<fe>
.
1
Connected frontend
Parameters:
<IPAddress>
string in double quotes
IP address of the frontend
<Subnet>
string in double quotes
Subnet mask of the frontend
<DHCP_State>
ON | OFF | 0 | 1
Indicates whether a DHCP server is used.
OFF | 0
DHCP off
ON | 1
DHCP on
*RST:
0
Example:
EFR:NETW?
//Result: "123.456.78.90","255.255.255.0",ON
Manual operation:
See
See
See
See
[SENSe:]EFRontend[:STATe]
<State>
Activates or deactivates the general use of an external frontend for the application.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
The frontend is disconnected. The application adapts the mea-
surement settings to the common settings supported by the
R&S
FSV/A.
ON | 1
The R&S
FSV/A allows you to configure and connect an external
frontend for the application. The application adapts the available
measurement settings to the connected frontend. The channel
bar indicates "Inp: ExtFe".
*RST:
0
Example:
EFR ON
Setting basic measurement parameters