![Rohde & Schwarz FSV3000 User Manual Download Page 785](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843785.webp)
Network operation and remote control
R&S
®
FSVA3000/ R&S
®
FSV3000
785
User Manual 1178.8520.02 ─ 08
11.2.2.8
STATus:QUEStionable:EXTended:INFO register
The
STATus:QUEStionable:EXTended:INFO
register contains information on the
type of messages that occur during operation of the R&S
FSV/A. A separate
INFO
reg-
ister exists for each active channel.
You can read out the register with
STATus:QUEStionable:EXTended:INFO:
or
STATus:QUEStionable:EXTended:INFO[:EVENt]?
. You can
query all messages that occur for a specific channel using the command
Table 11-8: Meaning of the bits used in the STATus:QUEStionable:EXTended:INFO register
Bit No.
Meaning
0
MESSage
This bit is set if event or state has occurred that may lead to an error during further operation.
1
INFO
This bit is set if an informational status message is available for the application.
2
WARNing
This bit is set if an irregular situation occurs during measurement, e.g. the settings no longer
match the displayed results, or the connection to an external device was interrupted temporarily.
3
ERRor
This bit is set if an error occurs during a measurement, e.g. due to missing data or wrong set-
tings, so that the measurement cannot be completed correctly.
4
FATal
This bit is set if a serious error occurs in the application and regular operation is no longer possi-
ble.
5 to 14
Unused
15
This bit is always 0.
11.2.2.9
STATus:QUEStionable:FREQuency register
The STATus:QUEStionable:FREQuency register contains information about the condi-
tion of the local oscillator and the reference frequency. A separate frequency register
exists for each active channel.
You can read out the register with
STATus:QUEStionable:FREQuency:
or
STATus:QUEStionable:FREQuency[:EVENt]?
Table 11-9: Meaning of the bits used in the STATus:QUEStionable:FREQuency register
Bit No.
Meaning
0
OVEN COLD
This bit is set if the reference oscillator has not yet attained its operating temperature. "OCXO"
is displayed.
1
LO UNLocked
This bit is set if the local oscillator no longer locks. "LOUNL" is displayed.
2 to 7
Not used
Status reporting system