Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
867
User Manual 1178.8520.02 ─ 08
12.5.3.1
Managing measurement configurations
The following commands control measurement configurations for ACLR measure-
ments.
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
............................................. 867
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
........................... 867
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete
.............................. 868
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE
.................................868
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
<Standard>
This command loads a measurement configuration.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Parameters:
<Standard>
For more information see
Chapter 5.3.9, "Reference: predefined
If you want to load a customized configuration, the parameter is
a string containing the file name.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
This command queries all files containing ACLR standards.
Suffix:
<n>
.
<m>
<sb>
Multi-SEM: 1 to 8
for all other measurements: irrelevant
Return values:
<Standards>
List of standard files.
Usage:
Query only
Manual operation:
See
Configuring and performing measurements