General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
723
User Manual 1178.8520.02 ─ 08
d) Repeat this to insert all other data points.
To insert a data point before an existing one, select the data point and then
"Insert Value".
To insert a new data point at the end of the list, move the focus to the line after
the last entry and then select "Insert Value".
To delete a data point, select the entry and then "Delete Value".
5. Check the current line configuration in the preview area of the dialog box. If neces-
sary, correct individual data points or add or delete some.
If necessary, shift the entire line vertically or horizontally by selecting the "Shift x" or
"Shift y" button and defining the shift width.
6. Save the new configuration by selecting the "Save" button.
The new transducer line is displayed in the overview and can be activated.
How to move the transducer line vertically or horizontally
A configured transducer line can easily be moved vertically or horizontally. Thus, a new
transducer line can be easily generated based upon an existing transducer line which
has been shifted.
1. In the "Line Config" dialog box, select the transducer line.
2. Select the "Edit" button.
3. In the "Edit Transducer Line" dialog box, select the "Shift x" or "Shift y" button and
define the shift width.
4. Save the shifted data points by selecting the "Save" button.
If activated, the trace is recalculated after the next sweep.
How to export a transducer factor
Transducer factor configurations can be stored to an ASCII file for evaluation in other
programs or to be imported later for other measurements.
1. In the "Edit Transducer" dialog box, select the transducer factor.
2. Select the "New" or "Edit" button.
3. Define the transducer factor as described in
"How to configure a new transducer
4. Select "Export" to save the configuration to a file.
You are asked whether you would like to save the configuration internally on the
R&S
FSV/A first.
5. Select a file name and location for the transducer factor.
6. Select the decimal separator to be used in the file.
Transducers