Message
Description
"Ext. Generator Command Error!"
Missing or wrong command in the generator setup file
(see
"Ext. Generator Visa Error!"
Error with Visa driver provided with installation (very
unlikely)
Overloading
At a reference level of -10 dBm and at an external generator output level of the same
value, the R&S
FSV/A operates without overrange reserve. That means the
R&S
FSV/A is in danger of being overloaded if a signal is applied whose amplitude is
higher than the reference line. In this case, either the message "RF OVLD" for over-
load or "ADC OVLD" for exceeded display range (clipping of the trace at the upper dia-
gram border = overrange) is displayed in the status line.
Overloading can be avoided as follows:
●
Reducing the output level of the external generator (
in "External Generator > Measurement Configuration")
●
Increasing the reference level (
6.2.4.3
External generator control settings
Access
: [INPUT/OUPUT] > "External Generator Config"
The "External Generator" settings are available if the R&S
FSV/A External Generator
Control option is installed. For each measurement channel, you can configure one
external generator. To switch between different configurations, define multiple mea-
surement channels.
For more information on external generator control, see
Interface configuration settings
............................................................................. 381
...........................................................................................383
..................................................................................386
Interface configuration settings
Access
: [INPUT/OUPUT] > "External Generator Config" > "Interface Configuration" tab
Data input and output