Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1161
User Manual 1178.8520.02 ─ 08
Example:
CALC:MATH:STAT ON
Turns on trace mathematics.
Manual operation:
See
See
12.8.2.4
Retrieving trace results
This chapter describes how to retrieve data from standard traces.
For spectrograms see also
Chapter 12.8.3.6, "Marker search (spectrograms)"
For details on the format of the retrieved trace data see also
mats for returned values: ASCII format and binary format"
Commands required to export traces (and other result data) are described in
ter 12.9.5, "Storing measurement results"
.........................................................................................1162
FORMat[:DATA]
<Format>[, <BitLength>]
This command selects the data format that is used for transmission of trace data from
the R&S
FSV/A to the controlling computer.
Note that the command has no effect for data that you send to the R&S
FSV/A. The
R&S
FSV/A automatically recognizes the data it receives, regardless of the format.
For details on data formats, see
Chapter 12.8.2.5, "Formats for returned values: ASCII
Parameters:
<Format>
ASCii
ASCii format, separated by commas.
This format is almost always suitable, regardless of the actual
data format. However, the data is not as compact as other for-
mats can be.
REAL
Floating-point numbers (according to IEEE 754) in the "definite
length block format".
<BitLength>
Length in bits for floating-point results
16
16-bit floating-point numbers.
Compared to
REAL,32
format, half as many numbers are
returned.
Analyzing measurements (basics)