Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1238
User Manual 1178.8520.02 ─ 08
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:STATe
<State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:THReshold
<Threshold>
This command defines a threshold for relative limit lines.
The R&S
FSV/A uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Threshold>
Numeric value.
The unit depends on
on page 1239.
*RST:
-200 dBm
Default unit: DBM
Manual operation:
See
Analyzing measurements (basics)