Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1109
User Manual 1178.8520.02 ─ 08
Note that the ratio defined with this remote command (RBW/span) is reciprocal to that
of the coupling ratio (span/RBW).
Parameters:
<Ratio>
Range:
0.0001 to 1
*RST:
0.01
Example:
BAND:RAT 0.1
Manual operation:
See
[SENSe:]BWIDth[:RESolution]:TYPE
<FilterType>
[SENSe:]BANDwidth[:RESolution]:TYPE
<FilterType>
This command selects the resolution filter type.
When you change the filter type, the command selects the next larger filter bandwidth if
the same bandwidth is unavailable for that filter.
The EMI-specific filter types are available if the EMI (R&S
FSV3-K54) measurement
option is installed, even if EMI measurement is not active. For details see
ter 5.13.3.1, "Resolution bandwidth and filter types"
Parameters:
<FilterType>
CFILter
Channel filters
NORMal
Gaussian filters
P5
5-pole filters
The 5-pole filter is not available for FFT sweeps.
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FSV3-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FSV3-K54) option
*RST:
NORMal
Example:
BAND:TYPE NORM
Example:
See
Chapter 12.15.1, "Programming example: performing a
Manual operation:
See
See
See
Setting basic measurement parameters