Figure 5-35: Adjacent channel power measurement on a CDMA2000 signal
5.3.7.2
Measurement example 2 – measuring adjacent channel power of a W-CDMA
uplink signal
Test setup:
Signal generator settings (e.g. R&S SMW):
Frequency:
1950 MHz
Level:
4 dBm
Modulation:
3GPP W-CDMA Reverse Link
Procedure:
1. Preset the R&S
FSV/A.
2. Set the center frequency to
1950
MHz
.
3. Select the "Channel Power ACLR" measurement function from the "Select Mea-
surement" dialog box.
4. Set the "W-CDMA 3GPP REV" standard for adjacent channel power measurement
in the "ACLR Setup" dialog box.
Channel power and adjacent-channel power (ACLR) measurement