Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1124
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
<w>
subwindow
Not supported by all applications
<t>
irrelevant
Parameters:
<Value>
numeric value WITHOUT UNIT (unit according to the result dis-
play)
Defines the range per division (total range = 10*<Value>)
*RST:
depends on the result display
Default unit: DBM
Example:
DISP:TRAC:Y:PDIV 10
Sets the grid spacing to 10 units (e.g. dB) per division
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
FSV/A adjusts the scaling of the y-axis accordingly.
For measurements with the optional external generator control, the command defines
the position of the reference value.
Suffix:
<n>
.
<w>
subwindow
Not supported by all applications
<t>
irrelevant
Parameters:
<Position>
0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST:
100 PCT = frequency display; 50 PCT = time dis-
play
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation:
See
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
Setting basic measurement parameters