Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
495
User Manual 1178.8520.02 ─ 08
Gating is not available for measurements on I/Q-based data.
Gated Trigger
Switches gated triggering on or off.
If the gate is switched on, a gate signal applied to one of the TRIGGER INPUT connec-
tors or the internal IF power detector controls the sweep.
Note:
For gated measurements, the maximum RBW is restricted to 3
MHz. If the cur-
rent RBW is larger when you enable gating, the RBW is automatically reduced to
3
MHz. See
For gated measurements in frequency sweep mode, only external triggers and EDGE
mode are supported, see
Remote command:
Gate Mode
Sets the gate mode.
For gated measurements in sweep mode, only the "Gate Mode"
Edge
is supported.
For more information see
Chapter 6.6.2.1, "Gated measurements"
"Edge"
The trigger event for the gate to open is the detection of the signal
edge.
After the gate signal has been detected, the gate remains open until
the gate length is over.
"Level"
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open until
the signal disappears.
Note:
If you perform gated measurements in combination with the IF
Power trigger, the R&S
FSV/A ignores the holding time for frequency
sweep, FFT sweep, zero span and I/Q mode measurements.
Remote command:
Gate Delay
Defines the delay time between the gate signal and the continuation of the measure-
ment.
For more information see
Chapter 6.6.2.1, "Gated measurements"
Remote command:
Gate Length
Defines how long the gate is open when it is triggered.