Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1202
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation:
See
12.8.3.8
Marker peak lists
Useful commands for peak lists described elsewhere
●
CALCulate<n>:MARKer<m>:PEXCursion
●
●
Chapter 12.8.3.3, "Configuring and performing a marker search"
Remote commands exclusive to peak lists
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt?
.................................................1203
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate]
............................................ 1203
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
...............................................1204
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT
.................................................... 1204
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STATe
....................................................1204
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?
......................................................... 1205
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
......................................................... 1205
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
<State>
This command turns labels for peaks found during a peak search on and off.
The labels correspond to the marker number in the marker peak list.
Suffix:
<n>
.
<m>
Analyzing measurements (basics)