![Rohde & Schwarz FSV3000 User Manual Download Page 959](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843959.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
959
User Manual 1178.8520.02 ─ 08
CALCulate<n>:ESPectrum:PSEarch:PSHow
<State>
CALCulate<n>:ESPectrum:PEAKsearch:PSHow
<State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:ESP:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation:
See
12.5.6.9
Performing an SEM measurement
The following commands are required to perform an SEM measurement:
●
SENS:SWE:MODE ESP
●
12.5.6.10
Retrieving results
The following commands analyze and retrieve measurement results for SEM measure-
ments.
●
●
●
●
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
12.5.6.11
Example: SEM measurement
In this example we will configure and perform an SEM measurement. Note that this
example is primarily meant to demonstrate the remote control commands, it does not
necessarily reflect a useful measurement task. For most common measurement stand-
ards, the R&S
FSV/A performs the measurement optimally with the predefined set-
tings, without further configuration.
//-----------Preparing the measurement ------------
//Reset the instrument
*RST
//------------ Preparing the measurement------------
//Activate SEM Measurement
Configuring and performing measurements