Storing the normalized reference trace as a transducer factor
The (inverse) normalized reference trace can also be stored as a
transducer factor
for
use in other R&S
FSV/A applications that do not support external generator control.
The normalized trace data is converted to a transducer with unit dB and stored in a file
with the specified name and the suffix
.trd
under
c:\r_s\instr\trd
. The fre-
quency points are allocated in equidistant steps between the start and stop frequency.
This is useful, for example, to determine the effects of a particular device component
and then remove these effects from a subsequent measurement which includes this
component.
For an example see
"How to remove the effects of a particular component from mea-
surement results using calibration"
Note that the
normalized
measurement data is stored, not the original
reference
trace!
Thus, if you store the normalized trace directly after calibration, without changing any
settings, the transducer factor will be 0
dB for the entire span (by definition of the nor-
malized trace).
Reference trace, reference line and reference level
Reference trace
The calibration results are stored internally on the R&S
FSV/A as a
reference trace
.
For each measured sweep point the offset to the expected values is determined. If nor-
malization is activated, the offsets in the reference trace are removed from the current
measurement results to compensate for the inherent distortions.
Reference line
The reference line is defined by the
and
in the
"External Generator" > "Source Calibration" settings. It is similar to the
defined in the "Amplitude" settings. However, as opposed to the reference
level
,
this reference
line
only affects the y-axis scaling in the diagram, it has no effect on the
expected input power level or the hardware settings.
The reference line determines the range and the scaling of the y-axis, just as the refer-
ence level does.
The normalized reference trace (0
dB directly after calibration) is displayed on this ref-
erence line, indicated by a red line in the diagram. By default, the reference line is dis-
played at the top of the diagram. If you shift the reference line, the normalized trace is
shifted, as well.
Shifting the reference line (and normalized trace)
You can shift the reference line - and thus the normalized trace - in the result display by
changing the
.
If the DUT inserts a gain or an attenuation in the measurement, this effect can be
reflected in the result display on the R&S
FSV/A. To reflect a power offset in the mea-
surement trace, change the
Data input and output