Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1134
User Manual 1178.8520.02 ─ 08
[SENSe:]SWEep:EGATe:TYPE
<Type>
This command selects the way gated measurements are triggered.
Parameters:
<Type>
LEVel
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open
until the signal disappears.
Note:
If you perform gated measurements in combination with
the IF Power trigger, the R&S
FSV/A ignores the holding time for
frequency sweep, FFT sweep, zero span and I/Q mode mea-
surements.
EDGE
The trigger event for the gate to open is the detection of the sig-
nal edge.
After the gate signal has been detected, the gate remains open
until the gate length is over.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation:
See
12.7.5.3
Configuring the trigger output
The following commands are required to send the trigger signal to one of the variable
"TRIGGER INPUT/OUTPUT" connectors on the R&S
FSV/A.
OUTPut<up>:TRIGger<tp>:DIRection
.............................................................................1134
.................................................................................. 1135
................................................................................ 1135
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
................................................................ 1136
OUTPut<up>:TRIGger<tp>:PULSe:LENGth
.................................................................... 1136
OUTPut<up>:TRIGger<tp>:DIRection
<Direction>
This command selects the trigger direction for trigger ports that serve as an input as
well as an output.
Suffix:
<up>
.
irrelevant
<tp>
Parameters:
<Direction>
INPut | OUTPut
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Setting basic measurement parameters