Common analysis and display functions
R&S
®
FSVA3000/ R&S
®
FSV3000
587
User Manual 1178.8520.02 ─ 08
Predefined Trace Settings - Quick Config
................................................................... 589
Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)
.............................................................590
Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the corresponding trace for configuration. The currently selected trace is high-
lighted.
Remote command:
Selected via numeric suffix of:
TRACe<1...6>
commands
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
Trace Mode
Defines the update mode for subsequent traces.
For details, see
"Analyzing several traces - trace mode"
"Clear/ Write"
Overwrite mode (default): the trace is overwritten by each sweep.
The "Detector" is automatically set to "Auto Peak".
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
FSV/A saves each trace point in the trace memory
only if the new value is greater than the previous one.
The "Detector" is automatically set to "PositivePeak".
This mode is not available for statistics measurements.
"Min Hold"
The minimum value is determined from several measurements and
displayed. The R&S
FSV/A saves each trace point in the trace mem-
ory only if the new value is lower than the previous one.
The "Detector" is automatically set to "Negative Peak".
This mode is not available for statistics measurements.
"Average"
The average is formed over several sweeps.
The
determines the number of averaging pro-
cedures.
The "Detector" is automatically set to "Sample".
This mode is not available for statistics measurements.
"View"
The current contents of the trace memory are frozen and displayed.
"Blank"
Removes the selected trace from the display.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE
Detector
Defines the trace detector to be used for trace analysis.
For details see
"Mapping samples to sweep points with the trace detector"
Trace configuration