Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
900
User Manual 1178.8520.02 ─ 08
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CALC:LIM:ACP:GAP2:MAN:LOW:ACLR:STAT BC,ON
Example:
See
"Example: configuring and performing an asymmetrical
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer[:CACLr][:
RELative]
<SBGaps>, <Limit>
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer[:CACLr][:
RELative]?
<SBGaps>
This command defines the relative limit of the specified lower gap (CACLR) channel.
The reference value for the relative limit is the measured channel power.
If you define both an absolute limit and a relative limit, the R&S
FSV/A uses the lower
value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the relative limit of the specified gap channel in dB.
Default unit: DB
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CALC:LIM:ACPower:GAP2:MANual:LOWer BC, 5
Example:
See
"Example: configuring and performing an asymmetrical
Configuring and performing measurements