Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1062
User Manual 1178.8520.02 ─ 08
[SENSe:]MIXer<x>:LOSS:HIGH
<Average>
This command defines the average conversion loss to be used for the entire high (sec-
ond) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<Average>
Range:
0 to 100
*RST:
24.0 dB
Default unit: dB
Example:
MIX:LOSS:HIGH 20dB
Manual operation:
See
[SENSe:]MIXer<x>:LOSS:TABLe:HIGH
<FileName>
This command defines the conversion loss table to be used for the high (second)
range.
Suffix:
<x>
.
1..n
irrelevant
Setting parameters:
<FileName>
String containing the path and name of the file, or the serial
number of the external mixer whose file is required. The
R&S
FSV/A automatically selects the correct cvl file for the cur-
rent IF. As an alternative, you can also select a user-defined
conversion loss table (
.acl
file).
Return values:
<FileName>
As the result of a query, the actually used file is returned.
Manual operation:
See
[SENSe:]MIXer<x>:LOSS:TABLe[:LOW]
<FileName>
This command defines the file name of the conversion loss table to be used for the low
(first) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<FileName>
String containing the path and name of the file, or the serial
number of the external mixer whose file is required. The
R&S
FSV/A automatically selects the correct cvl file for the cur-
rent IF. As an alternative, you can also select a user-defined
conversion loss table (
.acl
file).
Setting basic measurement parameters