Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1100
User Manual 1178.8520.02 ─ 08
Manual operation:
See
See
SYSTem:SPEaker:MAXVolume
<Volume>
Defines the maximum volume to be output as a percentage of the maximum possible
volume.
Parameters:
<Volume>
percentage
Example:
SYST:SPE:MAXV 50
Manual operation:
See
SYSTem:SPEaker:MUTE
Temporarily disables the audio output via the built-in loudspeakers.
Example:
SYST:SPE:MUTE
Manual operation:
See
SYSTem:SPEaker:VOLume
<Volume>
This command defines the volume of the built-in loudspeaker for demodulated signals.
This setting is maintained for all applications.
The command is available in the time domain in Spectrum mode and in Analog Modu-
lation Analysis mode.
Parameters:
<Volume>
Percentage of the maximum possible volume.
Range:
0 to 1
*RST:
0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
Manual operation:
See
12.7.2
Defining the frequency and span
The commands required to configure the frequency and span settings in a remote envi-
ronment are described here. The tasks for manual operation are described in
ter 6.3, "Frequency and span configuration"
.............................................................................. 1101
.................................................................................. 1105
Setting basic measurement parameters