Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1057
User Manual 1178.8520.02 ─ 08
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<State>
OFF | ON | AUTO | ALL
OFF | ON | AUTO | ALL
OFF
No automatic signal detection is active.
ON
Automatic signal detection (Signal ID) is active.
AUTO
Automatic signal detection (Auto ID) is active.
ALL
Both automatic signal detection functions (Signal ID+Auto ID)
are active.
*RST:
OFF
Manual operation:
See
See
[SENSe:]MIXer<x>:THReshold
<Value>
This command defines the maximum permissible level difference between test sweep
and reference sweep to be corrected during automatic comparison (see
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<Value>
<numeric value>
Range:
0.1 dB to 100 dB
*RST:
10 dB
Default unit: DB
Example:
MIX:PORT 3
Manual operation:
See
Mixer settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer<x>:FREQuency:HANDover
.................................................................... 1058
[SENSe:]MIXer<x>:FREQuency:STARt
.......................................................................... 1058
[SENSe:]MIXer<x>:FREQuency:STOP
...........................................................................1058
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
................................................................ 1059
[SENSe:]MIXer<x>:HARMonic:BAND
.............................................................................1059
Setting basic measurement parameters