Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
461
User Manual 1178.8520.02 ─ 08
Or
: [AMPT] > "Scale Config"
The remote commands required to define these settings are described in
ter 12.7.4, "Configuring the vertical axis (amplitude, scaling)"
Range
Defines the displayed y-axis range in dB.
The default value is 100
dB.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]
Ref Level Position
Defines the reference level position, i.e. the position of the maximum AD converter
value on the level axis in %.
0
% corresponds to the lower and 100
% to the upper limit of the diagram.
Values from -120
% to +600
% are available. Larger values are useful for small scales,
such as a power range of 10
dB or 20
dB, and low signal levels, for example 60
dB
below the reference level. In this case, large reference level position values allow you
to see the trace again.
For spectrograms, this value defines the position of the reference level value within the
span covered by the color map. In this case, the value is given in %, where 0
% corre-
sponds to the maximum (right end) and 100
% to the minimum (left end) of the color
map.
Amplitude and vertical axis configuration