Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
403
User Manual 1178.8520.02 ─ 08
B600/B1000-specific phase conversion loss tables are provided in files according to
the following syntax:
<serial_no.>_PHASE_<harmonic>_B2000.b2g
Automatic signal identification
Automatic signal identification allows you to compare the upper and lower band results
of the mixer, thus detecting unwanted mixer products due to conversion.
Note that automatic signal identification is only available for measurements that per-
form frequency sweeps (not in vector signal analysis or the I/Q Analyzer, for instance).
The "Auto ID" function is now also available for
Spectrum emission mask (SEM) mea-
Spurious emissions measurement
Signal ID function
Two sweeps are performed alternately. Trace 1 shows the trace measured on the
upper side band (USB) of the LO (the test sweep), trace 2 shows the trace measured
on the lower side band (LSB), i.e. the reference sweep.
Figure 6-16: Signal identification function (Signal ID) with optional external mixer
The reference sweep is performed using an LO setting shifted downwards by 2*IF/
<Harmonic order>. Input signals in the desired sideband that are converted using the
Data input and output