Applications, measurement channels
R&S
®
FSVA3000/ R&S
®
FSV3000
130
User Manual 1178.8520.02 ─ 08
Example: Sequencer procedure
Assume the following active channel definition:
Tab name
Application
Sweep mode
Sweep count
Spectrum
Spectrum
Cont. Sweep
5
Spectrum 2
Spectrum
Single Sweep
6
Spectrum 3
Spectrum
Cont. Sweep
2
IQ Analyzer
IQ Analyzer
Single Sweep
7
For
Single Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer
For
Continuous Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
...
For
Channel-defined Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 2 x Spectrum 3,
5x Spectrum, 2 x Spectrum 3,
...
Run Single/Run Cont and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the [Run Single] and [Run Cont] keys control the
Sequencer, not individual sweeps. [Run Single] starts the Sequencer in single mode,
while [Run Cont] starts the Sequencer in continuous mode.
Running a sequence of measurements