Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1242
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:UPPer:STATe
<State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:THReshold
<Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S
FSV/A uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Limit>
Numeric value.
The unit depends on
on page 1239.
*RST:
-200
Default unit: dBm
Manual operation:
See
12.8.5.2
Managing limit lines
Useful commands for managing limit lines described in the R&S
FSV/A User Manual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
●
MMEM:LOAD:TYPE
Analyzing measurements (basics)