Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
343
User Manual 1178.8520.02 ─ 08
Filter bandwidth:
[SENSe:]BANDwidth[:RESolution]
5.13.4.3
LISN control settings
Access
: [MEAS CONFIG] > "LISN Config"
For measurements with power lines, the following settings are available for the
R&S
FSV/A to control which phase of the LISN is to be tested (e.g. for EMI measure-
ments). LISN control requires the EMI measurement (R&S
FSV3-K54) option.
For more information, see
Chapter 5.13.3.4, "Controlling V-Networks (LISN)"
LISN Type
Selects the network type and activates output to the network via the user port of the
R&S
FSV/A. The network type determines the supported phases (see
).
"Off" disables LISN control and output.
Remote command:
Phase
Selects the phase to be measured. Phase N and L1 are included in all four LISN.
Phase L2 and L3 are only included in four-line networks.
You can select one phase only for each measurement.
Electromagnetic interference (EMI) measurement