![Rohde & Schwarz FSV3000 User Manual Download Page 174](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843174.webp)
Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
174
User Manual 1178.8520.02 ─ 08
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
Span: determined signal bandwidth plus 40
%
●
TX and adjacent channels
●
ACLR standard according to estimated signal bandwidth
●
For pulsed signals: gating parameters
After the parameters have been determined automatically, you must confirm the ACLR
standard selection.
See also
Chapter 5.1, "Auto measurements"
The remote commands required to perform these tasks are described in
ter 12.5.3, "Measuring the channel power and ACLR"
General CP/ACLR measurement settings
............................................................ 174
5.3.4.1
General CP/ACLR measurement settings
General measurement settings are defined in the "ACLR Setup" dialog, in the "General
Settings" tab.
................................................................................... 175
............................................................................................. 175
.......................................................................................176
Absolute and Relative Values (ACLR Mode)
.............................................................. 177
Channel power level and density (Power Unit)
........................................................... 177
Setting a fixed reference for Channel Power measurements (Set CP Reference)
Optimized Settings (Adjust Settings)
...........................................................................178
Standard
The main measurement settings can be stored as a standard file. When such a stan-
dard is loaded, the required channel and general measurement settings are automati-
cally set on the R&S
FSV/A. However, the settings can be changed. Predefined stand-
ards are available for standard measurements, but standard files with user-defined
configurations can also be created.
Channel power and adjacent-channel power (ACLR) measurement