Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1150
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
1..n
Parameters:
<Mode>
LOGarithmic
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
For details see
"Mapping samples to sweep points with the trace detector"
For EMI measurements, the trace detector is used for the initial peak search only, not
for the final test. The detector for the final test is configured using
MARKer<m>:FUNCtion:FMEasurement:DETector
If the EMI (R&S
FSV3-K54) measurement option is installed and the filter type "CISPR"
is selected, additional detectors are available, even if EMI measurement is not active.
For details see
Chapter 5.13.3.2, "Detectors and dwell time"
Suffix:
<n>
.
<t>
Example:
DET POS
Sets the detector to "positive peak".
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
<t>