![Rohde & Schwarz FSV3000 User Manual Download Page 373](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843373.webp)
Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
373
User Manual 1178.8520.02 ─ 08
Signal generator
rear panel
Signal analyzer
rear panel
BNC Blank
BNC Trigger
Figure 6-5: TTL connection for an R&S SMA100B generator
The external generator can be used to calibrate the data source by performing either
transmission or reflection measurements.
Transmission Measurement
This measurement yields the transmission characteristics of a two-port network. The
external generator is used as a signal source. It is connected to the input connector of
the DUT. The input of the R&S
FSV/A is fed from the output of the DUT. A calibration
can be carried out to compensate for the effects of the test setup (e.g. frequency
response of connecting cables).
Figure 6-6: Test setup for transmission measurement
Reflection Measurement
Scalar reflection measurements can be carried out using a reflection-coefficient mea-
surement bridge.
Figure 6-7: Test setup for reflection measurement