General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
756
User Manual 1178.8520.02 ─ 08
Frequency ← Calibration Frequency RF
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration (max. 64
MHz).
For narrowband signals, 64
MHz is sent.
Calibration Frequency MW
Uses the microwave calibration signal as the RF input (for frequencies higher than 8
GHz). This function is used to calibrate the YIG-filter on the microwave converter. The
microwave calibration signal is pulsed.
You can define the frequency of the microwave calibration signal. The firmware adapts
the defined value to the next possible valid value.
Note:
The microwave calibration signal is only displayed if the
Chapter 6.2.2.1, "Radio frequency input"
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:MC:CFRequency
10.9.4
Service functions
Access
: [Setup] > "Service" > "Service Function"
Using service functions
The service functions are not necessary for normal measurement operation. Incorrect
use can affect correct operation and/or data integrity of the R&S
FSV/A.
Therefore, only user accounts with administrator rights can use service functions and
many of the functions can only be used after entering a password. These functions are
described in the instrument service manual.
Service functions