Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1010
User Manual 1178.8520.02 ─ 08
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
//Turn on high pass filter
INP:LISN:TYPE ENV216
INP:LISN:FILT:HPAS ON
Manual operation:
See
INPut<ip>:LISN:PHASe
<Phase>
This command selects one LISN phase to be measured.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Phase>
L1
L2
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
L3
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
N
*RST:
L1
Example:
//Select phase L1
INP:LISN:PHAS L1
Manual operation:
See
INPut<ip>:LISN[:TYPE]
<Type>
This command turns automatic control of a LISN on and off. It also selects the type of
network.
Configuring and performing measurements