Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Auto Detector (Auto Peak)
Traces 2-6:
Blank
Auto Detector
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Auto Detector (Positive Peak)
Trace 2:
Average
Auto Detector (Sample)
Trace 3:
Min Hold
Auto Detector (Negative Peak)
Traces 4-6:
Blank
Auto Detector
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Auto Detector (Positive Peak)
Trace 2:
Clear Write
Auto Detector (Auto Peak)
Trace 3:
Min Hold
Auto Detector (Negative Peak)
Traces 4-6:
Blank
Auto Detector
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:PRESet
Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
For details see
Chapter 7.5.1.3, "How to configure a standard trace"
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
Copy Trace
Access
: "Overview" > "Analysis" > "Traces" > "Copy Trace"
Copies trace data to another trace.
The first group of buttons (labeled "Trace 1" to "Trace 6") selects the source trace. The
second group of buttons (labeled "Copy to Trace 1" to "Copy to Tace 6") selects the
destination.
Remote command:
Trace Labels
Access
: "Overview" > "Analysis" > "Traces" > "Trace Label" tab
Trace configuration