CALCulate<n>:LIMit<li>:ESPectrum<sb>:LIMits
<Max1>,<Max2>,<Max3>
This command sets or queries up to 4 power classes in one step. You can only define
values for the number of power classes defined by
ESPectrum<sb>:PCLass<pc>:COUNt
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 8
Setting parameters:
<Max1>
Defines the value range for power class 1 as
-200 to
<Max1>
.
Only available for
CALC:LIM:ESP:PCL:COUNT >=2
If only 2 power classes are defined, the value range for power
class 2 is defined as
<Max1> to 200
.
Range:
-199 to + 199
Default unit: DBM
<Max2>
Defines the value range for power class 2 as
<Max1> to
<Max2>
.
Only available for
CALC:LIM:ESP:PCL:COUNT >=3
If only 3 power classes are defined, the value range for power
class 3 is defined as
<Max2> to 200
.
Range:
-199 to + 199, <Max2> must be higher than
<Max1>
<Max3>
Defines the value range for power class 3 as
<Max2> to
<Max3>
.
The value range for power class 4 is defined as
<Max3> to
200
.
Only available for
CALC:LIM:ESP:PCL:COUNT = 4
Range:
-199 to + 199, <Max3> must be higher than
<Max2>
Example:
CALC:LIM:ESP:LIM -50,50,70
Defines the following power classes:
<-200, -50>
<-50, 50>
<50, 70>
<70, 200>
Query:
CALC:LIM:ESP:LIM?
Response:
-200,-50,50,70,200
Configuring and performing measurements