Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1098
User Manual 1178.8520.02 ─ 08
OUTPut<up>:IF:STATe
<State>
Enables or disables output of the measured IF value at the "IF" output connector.
Suffix:
<up>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
OUTP:IF:STAT ON
Manual operation:
See
OUTPut<up>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSV/A.
For the R&S
FSV/A, this command is maintained for compatibility reasons only. The
AUX PORT
connector only supports output.
Suffix:
<up>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
User port is switched to INPut
ON | 1
User port is switched to OUTPut
Example:
OUTP:UPOR:STAT ON
OUTPut<up>:UPORt[:VALue]
<Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit
7
6
5
4
3
2
1
0
Pin
N/A
N/A
5
3
4
7
6
2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Setting basic measurement parameters