![Rohde & Schwarz FSV3000 User Manual Download Page 1261](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431261.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1261
User Manual 1178.8520.02 ─ 08
It is therefore recommended that you use the appropriate
MMEMory:SELect[:ITEM]
command before using
MMEMory:LOAD:STATe
.
Parameters:
<1>
<FileName>
String containing the path and name of the file to load.
The string may or may not include the file's extension.
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S
FSV/A.
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument.
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
See
See
"Recall in New Channel / Recall in Current Channel"
MMEMory:LOAD:TYPE
<Type>
This command defines whether the channels that will be loaded with the subsequent
MMEM:LOAD:STAT
command will replace the current channel or activate a new chan-
nel.
Parameters:
<Type>
NEW | REPLace
NEW
The loaded settings will be activated in a new channel.
REPLace
The loaded settings will replace the currently active channel.
*RST:
NEW
Managing settings and results