
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
943
User Manual 1178.8520.02 ─ 08
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:RLEVel
<RefLevel>
This command defines the reference level for a SEM range.
In case of high speed measurements, the reference level has to be identical for all
ranges.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<RefLevel>
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Default unit: dBm
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME
<SweepTime>
This command defines the sweep time for a SEM range.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<SweepTime>
Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Default unit: s
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a SEM range on and
off.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Configuring and performing measurements