Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
191
User Manual 1178.8520.02 ─ 08
For symmetrical channel definition (see
the dialog box is reduced as the upper and lower channels are identical.
Figure 5-30: Asymmetrical adjacent channel definition
For details on setting up channels, see
Chapter 5.3.6.3, "How to configure an MSR
Number of Adjacent Channels (Adj Count)
................................................................. 191
........................................................................................192
..........................................................................192
...................................................................... 192
........................................................................................... 192
.............................................................................................. 193
Number of Adjacent Channels (Adj Count)
Defines the number of adjacent channels
above and below
the Tx channel block in an
MSR signal. You must define the carrier channel to which the relative adjacent-channel
power values refer (see
Remote command:
[SENSe:]POWer:ACHannel:ACPairs
Channel power and adjacent-channel power (ACLR) measurement