![Rohde & Schwarz FSV3000 User Manual Download Page 1074](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431074.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1074
User Manual 1178.8520.02 ─ 08
For details see
"Multimode function and offset compensation for modular RF probes"
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<Mode>
CM | DM | PM | NM
DM
Voltage between the positive and negative input terminal
CM
Mean voltage between the positive and negative input terminal
vs. ground
PM
Voltage between the positive input terminal and ground
NM
Voltage between the negative input terminal and ground
Example:
SENS:PROB:SETU:PMOD PM
Sets the probe to P-mode.
Manual operation:
See
[SENSe:]PROBe<pb>:SETup:PMOFfset
<PMOffset>
Sets the P-mode offset. The setting is only available if a modular probe in P-mode is
connected to the R&S
FSV/A. The maximum voltage difference between the positive
and negative input terminals is 16
V.
If the probe is disconnected, the P-mode offset of the probe is reset to 0.0
V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see
"Multimode function and offset compensation for modular RF probes"
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<PMOffset>
The voltage offset between the positive input terminal and
ground.
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
Setting basic measurement parameters